Visite nuestro sitio en Español: Confiabilidad.net    RSS | Contact

Oil Analysis – Circulating System

April 29, 2009
(Oil and Fluid Analysis)

Oil analysis testing can be expensive and sometimes performing every test on a sample is not required. When sampling a circulating system, only one sample should be fully evaluated: the primary sample. The primary sample should be taken from a return line if at all possible. If it is not possible to sample at the return line, then either the reservoir or before the filtration system would be the next best places. Consider all other samples secondary in nature. The primary sample should get all required testing such as Viscosity, Karl Fischer, FTIR, Spectroscopy, Particle Count, etc to evaluate the chemical and physical properties of the oil. The secondary samples should only have a particle count until an abnormal result warrants further testing. Rationalizing what tests are required on samples will lower the amount of information coming back to you and save you money

Tip provided by Kris Sonne
Predictive Maintenance Technician
Trico Corp and Predict Inc.

 

 

Previous tip: Selling a house and maintenance really do have a lot in common
Next tip: Electrical Stator faults

« Back to all maintenance tips

Comments

There are no comments for this article yet. Why not add one?

Have your say

Comments are moderated prior to publication.
Please fill out the fields below.
Email addresses will never be published.

Comment guidelines

You can use basic HTML (a, strong, em, blockquote).
Links automatically use the nofollow attribute.
Off-topic or inappropriate comments will be edited or deleted.

Advertisement